RECIRCULATING SYSTEM
The operation of a recirculating system is similar
to that of a single pass unit. With this system the
air is recirculated through a perforated work deck
to a rear-mounted return air duct. This returns the
air to the blower intake where it is forced through
the HEPA filter again and returned back to the clean
zone
SINGLE PASS SYSTEM
Ambient air enters the pre-filter area and is forced
through a HEPA filter, affording downward laminar flow
through the clean zone. The air is then exhausted out
the front of the unit. If necessary, the work deck
can be perforated to allow air to pass directly through
the unit and then out below the work deck.
CLEAN FUME HOOD-EXHAUST
The
design of the Clean Fume Hood-Exhaust includes a perforated
work deck. Ambient air enters the system at the top,
pressurizes the blower plenum and then the clean air
is forced through the HEPA filter with 99.99% effectiveness.
The air travels in a laminar flow down through the
perforated work surface into the plenum area below
and exhausts through the rear mounted exhaust duct
to the outside of the central exhaust system as determined
by the design.
VARIABLE EXHAUST
The Variable Exhaust unit affords the flexibility
of exhausting a portion of the returned air (0% to
100%) resulting in a savings of ambient air conditioning
and extending the life of the HEPA filter. The variable
exhaust system may be appropriate where a negligible
fume situation exists. |